Link FLEX VISION (ROM) devices with the computer by holding down the red button.
Select the section to be scanned.
Always make sure the devices are facing the same direction, and zero out devices on the Patient on each set.
Cervical Flexion/ Extension:
Hold the devices in place and then press the red button to zero out the devices.
Place the device with the white circle in your left hand and place it on the patients head with the front of the device facing you.
Place the other device in your right hand and place the probes at T1, T2.
Ask Patient to tilt their head forward (flexion). Wait (1 1,000) for the bar to stop moving, then press the button.
Ask Patient to return to a neutral position for a brief second.
Ask Patient to tilt their head back (extension). Wait (1 1,000) for the bar to stop moving, then press the button.
Ask Patient to return to a neutral position.
Repeat each motion 3 times total following the directions on the screen.
For each movement make sure the devices are stable, never remove the devices from the patient and make sure the neutral device (right) is not moving.
Cervical Left / Right Lateral:
Stand behind the patient.
Hold the devices in place and then press the red button to zero out the devices.
Place the device with the white circle in your left hand and place it on the patients head with the front of the device facing you.
Place the other device in your right hand and place the probes at T1, T2.
Ask the Patient to tilt their head all the way to the left (left lateral flexion). Wait (1 1,000) for the bar to stop moving, then press the button.
Ask the Patient to return to a natural position for a brief second.
Ask the Patient to tilt their head all the way to the right (right lateral flexion). Wait (1 1,000) for the bar to stop moving, then press the button.
Ask the Patient to return to a natural position.
Repeat each motion 3 times total following the directions on the screen.
For each movement make sure the devices are stable, never remove the devices from the patient and make sure the neutral device (right) is not moving.
Cervical Left / Right Rotation:
Ask the Patient sit on the table, and bend forward then zero out the devices. Press the red button.
The device with the white circle should be in your left hand and should be pressed against the patients back of the head.
The other device, needs to have the probes pressed firmly against the patients T1, T2 Rib.
Ask the Patient to look all the way to the left slowly (left rotation). Wait (1 1,000) for the bar to stop moving, then press the button.
Ask the Patient to return to a natural position for a brief second.
Ask the Patient to look all the way to the right slowly. Wait (1 1,000) for the bar to stop moving, then press the button.
Ask the Patient to return to a neutral position.
Repeat each motion 3 times total following the directions on the screen.
For each movement make sure the devices are stable, never remove the devices from the patient and make sure the neutral device (right) is not moving.
Lumbar Flexion/ Extension:
Ask the Patient to stand, place left device with white dot on T11 and right device on L5 level. Zero out devices on patient. Press the red button. (Men: bottom of scapula. Women: Bra line)
Ask the Patient to bend forward (lumbar flexion). Wait (1 1,000) for the bar to stop moving, then press the button.
Ask the Patient to stand back up in a natural position for a brief second.
Ask the Patient to bend backwards (lumbar extension). Wait (1 1,000) for the bar to stop moving, then press the button.
Repeat each motion 3 times total following the directions on the screen.
For each movement make sure the devices are stable, never remove the devices from the patient and make sure the neutral device (right) is not moving.
Lumbar Left/ Right Lateral:
Place left device with white dot on patients T11 and right device on L4, L5 location.
Ask the Patient to stand up nice and tall. Press the red button to zero out the devices.
Ask the Patient to slide their left arm down the side of their body (left lateral flexion). Wait (1 1,000) for the bar to stop moving, then press the button.
Ask the Patient to return to a neutral position for a brief second.
Ask the Patient to slide their right arm down the side of their body (right lateral flexion). Wait (1 1,000) for the bar to stop moving, then press the button.
Ask the Patient to return to a neutral position.
Repeat each motion 3 times total following the directions on the screen.
For each movement make sure the devices are stable, never remove the devices from the patient and make sure the neutral device (right) is not moving.
When the test is complete click on “Review” and the ROM pie graphs will appear.